GW Instek GDM-8245 User Manual

Page 6

Advertising
background image

DUAL DISPLAY DIGITAL MULTIMETER

USER MANUAL

8

Protection

500μA,5mA,50mA,500mA,2A 5 ranges fuse protection.
20A range no fuse,15 seconds max.

When the input exceeds the full scale of the selected range, the display will appear
“—OL—“ of over-range indication.
5. TRUE RMS AC OR AC+DC CURRENT

Accuracy

Between 2% of range and full range.

RANGE

20Hz-45Hz

45Hz-2kHz

2kHz-10kHz

10kHz-20kHz

500μA

1%+15

0.5%+15

1%+15

2%+15

5mA

1%+15

0.5%+15

1%+15

2%+15

50mA

1%+15

0.5%+15

1%+15

2%+15

500mA

1%+15

0.5%+15

—————————

2A

1%+15

0.5%+15

—————————

20A

1%+15

0.5%+15

—————————

Protection

500μA,5mA,50mA,500mA,2A 5 ranges fuse protection.
20A range no fuse,15 seconds max.

Crest Factor Range 3.0 at full scale.
The burden voltage is the same as the DC current.

When the input exceeds the full scale of the selected range, the display will appear
“—OL—” of over-range indication.

6. FREQUENCY MEASUREMENT AT ACA RANGE

RANGE FREQUENCY INPUT LEVEL (SINE WAVE)

ACCURACY

500μA 10Hz ~ 20kHz

≧

90μA

0.05%+1

5mA

10Hz ~ 20kHz

≧

0.9mA

0.05%+1

50mA

10Hz ~ 20kHz

≧

9mA

0.05%+1

500mA

10Hz ~ 20kHz

≧

90mA

0.05%+1

2A

10Hz ~ 2kHz

≧

1A

0.05%+1

20A

10Hz ~ 2kHz

≧

9A

0.05%+1

AC+DC measurement does not support AC+Hz function.

DUAL DISPLAY DIGITAL MULTIMETER

USER MANUAL

9

7.RESISTANCE

RANGE

RESOLUTION

ACCURACY

500Ω

0.01Ω

0.1%+4

5kΩ

0.1Ω

0.1%+2

50kΩ

1Ω

0.1%+2

500kΩ

10Ω

0.1%+2

5MΩ

100Ω

0.2%+2

20MΩ

1kΩ

0.3%+2

Open-circuit Voltage

3.2 volts maximum on 500Ω, 1.3 volts maximum on all other

ranges.

Protection

450V dc or peak ac continuous.

8. CAPACITANCE

RANGE:

RESOLUTION

ACCURACY

5n *

0.001n

≧1nF: 2%+10

<1nF & ≧0.5nF: 2%+20

50n

0.01n

≧10nF: 2%+10

<10nF & ≧5nF: 2%+30

500n

0.1n

2%+4

5μ

1n

2%+4

50μ

10n

2%+4

*5n range tends to be interfered by the test lead’s impedance and position. For the
accuracy, please measure the range on the input terminal directly.
Protection

450V dc or peak ac continuous.

9. DIODE CHECK

Description

Display read forward voltage of diode.

Open Voltage

3.1V approx.

Maximum Forward
Voltage

1.5V

Protection

450V dc or peak ac continuous.

Advertising