Resistance, Diode, Continuity – GW Instek GDM-8300 series User Manual User Manual
Page 152

GDM-834X Series User Manual
152
Resistance
Resistance
Resolution
Full Scale
Test Current
Accuracy
(1 year 23˚C
±5˚C)[2]
500Ω
10mΩ
510.00
0.83mA
0.1%+5 [1]
5kΩ
100mΩ
5.1000
0.83mA
0.1%+3 [1]
50kΩ
1Ω
51.000
83
μ
A
0.1%+3
500kΩ
10Ω
510.00
8.3
μ
A
0.1%+3
5MΩ
100Ω
5.1000
830nA
0.1%+3
50MΩ
1KΩ
51.000
560nA//10MΩ
0.3%+3
[1] Using the REL function. If you don't use the REL function then increase the
error by 0.2Ω.
[2] When measuring resistances greater than 500kΩ, please use shielded test leads
to eliminate the noise interference that may be induced by standard test leads.
* Open circuit voltage approximates 6V max on 500~5MΩ range, approximates
5.5V max on 50MΩ range.
* Input protection of 500V peak on all ranges.
Diode
Range
Resolution
Full Scale
Test Current
Accuracy
(1 year 23˚C ±5˚C)
5V
100
μ
V
5.1000
0.83mA
0.05%+5
* Input protection of 500V peak. *Open circuit voltage approximates 6V.
Continuity
Range
Resolution
Full Scale Test Current
Accuracy
(1 year 23˚C ±5˚C)
5000.0Ω
100mΩ
5100.0
0.83mA
0.1%+5
* Input protection of 500V peak. *Open circuit voltage approximates 6V.