SCANNER BOX HSB-001-1/2
USER MANUAL
10
EXAMPLE OF SETTING UP THE HSB-001-1/2
1:01 ACW V = 1.200 kV EDIT
Imax=03.00 mA TEST:060.0S
1:01 ACW V = 1.200 kV EDIT
Channel Hi= 01 Lo=02
When the function jumps from Filed to Channel, select Hi from
01~16 and Lo from 01~16, such as Hi=05 and Lo=01, means a high
voltage will be outputted from the Channel 05, and the channel 01
is a return terminal.
HSB-001-1 : ACW,DCW,IR for Channel 01~08, GB for Channel
09~16
HSB-001-2 : ACW,DCW,IR for Channel 01~16,
If select Channel=00 and Channel=00, there will be no output
from scanner box.
SCANNER BOX HSB-001-1/2
USER MANUAL
11
z
The Configuration
HSB-001-1
HSB-001-1 and GPI-745A