GDS-800 Series Digital Storage Oscilloscope Operation Manual
48
z
Previous Menu: back to previous menu.
For Template Auto
z
Template Auto: The two “Go-No Go” templates are generated
automatically from the subject signal.
(1): According to the subject signal, the two templates (RefA and RefB) are
created by “Template Auto” automatically. The adjustable range (Tolerance)
is available from ±0.4% to ±40% (or ±0.04 division to ±4 division).
Figure 5-23: Template created “Template Auto” function
z
Source CH1/CH2: Select the input source of channel 1 or channel 2 as
“Go-No Go” template.
z
Tolerance %: Select the tolerance percentage (or division) range for
vertical and horizontal scales from the subject signal. The adjustable range
is available from ±0.4% to ±40% (or ±0.04 division to ±4 division).
z
Save & Create: Press F4 softkey to save the settings. In the meantime,
the original saved reference A or B from “Save/Recall” function will be also
changed as current setting
Previous Menu: Back to previous menu.