Screwdriver portable calibrator – HT instruments HT154 User Manual

Page 9

Advertising
background image

HT154

EN - 8

4.3.2. Calibration

procedure

Screwdriver

Portable calibrator

HT154

Sound Level Meter

Fig. 4: Connection of meter to portable calibrator HT150


1. Turn ON the meter.
2. Setting herewith configuration:

Frequency Weighting: A (dBA)
Integration time: FAST

3. Inserting with care the portable calibrator HT150 on meter microphone up to listen a

click, as shown in Fig.4.

4. Pressing LEVEL key and selecting “50-100” range at display.
5. Turn ON calibrator and select 94dB level at 1kHz sinusoidal frequency. Red LED on

calibrator must be ON.

6. Insert the screwdriver inside “CAL” trimmer and adjust the measured value up to read

“94.0” at display.

7. Extract from microphone and switch OFF calibrator.