EN - 3

1.2. DURING

USE

Read the recommendations which follow and the instructions in this manual:

CAUTION

An improper use may damage the instrument and/or its components or injure
the operator.

• When changing the range, first disconnect the test leads from the circuit under test in

order to avoid any accident

• When the instrument is connected to measuring circuits never touch any unused

terminal

• When measuring resistors do not add any voltage. Although there is a protection

circuit, excessive voltage could cause malfunctioning

• If during measurement the displayed values remain constant check whether the HOLD

function is active

1.3. AFTER

USE

• After using the instrument turn it off
• If you expect not to use the instrument for a long period remove the battery to avoid

leakages of battery liquids which may damage its inner components

1.4. MEASURING

(OVERVOLTAGE) CATEGORIES DEFINITIONS

EN 61010-1: Safety requirements for electrical equipment for measurement, control and
laboratory use, Part 1: General requirements, gives a definition of measuring category,
usually called overvoltage category. Paragraph 6.7.4: Measuring circuits:

(OMISSIS)

circuits are divided into the following measurement categories:

• Measurement category IV is for measurements performed at the source of the low-

voltage installation

Examples are electricity meters and measurements on primary overcurrent protection
devices and ripple control units

• Measurement category III is for measurements performed in the building installation.

Examples are measurements on distribution boards, circuit breakers, wiring, including
cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and
equipment for industrial use and some other equipment, for example, stationary
motors with permanent connection to fixed installation

• Measurement category II is for measurements performed on circuits directly

connected to the low voltage installation

Examples are measurements on household appliances, portable tools and similar
equipment

• Measurement category I is for measurements performed on circuits not directly

connected to MAINS

Examples are measurements on circuits not derived from MAINS, and specially
protected (internal) MAINS-derived circuits. In the latter case, transient stresses are
variable; for that reason, the norm requires that the transient withstand capability of the
equipment is made known to the user