HT309

EN - 5

4.2. DESCRIPTION OF FUNCTION KEYS
Some of the function keys on the front panel have a double function, which can be
activated by pressing and holding the key for more than 1 second.

4.2.1. HOLD/L.S. (Light Source) key
¾

Press the HOLD/L.S key to freeze the value of the measured quantity on the display.

The symbol “HOLD” appears on the display. Press the HOLD/L.S. key again to exit the
function.


¾

Press and hold the HOLD/L.S. key for more than 1 second to select the light source to

be measured. Proceed as follows:


1. On the sub-display, observe the flashing symbol “Lx”, where 0 < x < 9 indicates the

source number.

2. Use the arrow keys S or T to select the desired source among the available options

L0 ÷ L9 (see Table 1)

3. Press the HOLD/L.S. key to set a possible correction factor on the instrument’s reading

(only for custom sources L7, L8 and L9). The indication of the parameter value flashes
on the display. Use the arrow keys S or T to set (press and hold the keys for rapid
operation) the value between 0.001 and 1.999. The values of the sources L0 ÷ L5
cannot be modified.

Symbol on the

display

Source type

Correction factor

L0 Standard

light

1.000

L1 White

LED

0.990

L2 Red

LED

0.516

L3 Yellow

LED

0.815

L4 Green

LED

1.216

L5 Blue

LED

1.475

L6 Purple

LED

1.148

L7 Custom

1.000

L8 Custom

1.000

L9 Custom

1.000

Table 1: List of available sources

4. Press and hold the HOLD/L.S. key for more than 1 second to confirm the operation.

The symbols stop flashing on the display.

4.2.2. LX/FC/CD

key

¾

Press the LX/FC/CD key to select the measuring unit of illuminance, choosing between

the two available options “LUX” and “FC” (1Fc = 10.764Lux ; 1Lux = 0.09290Fc).


¾

Press and hold the LX/FC/CD key for more than 1 second to select the measuring unit

of luminous intensity (see § 4.3.2).