Safety precautions and procedures – HT instruments HT327 User Manual

Page 3

Advertising
background image

HT327

EN - 2

1. SAFETY PRECAUTIONS AND PROCEDURES

This meter is in compliance with safety IEC/EN61010-1 guideline related to electronic
measuring instruments. For your own safety and to avoid damaging the instrument follow
the procedures described in this instruction manual and read carefully all notes preceded
by this symbol

.When taking measurements:


 Avoid doing that in humid or wet places - make sure that humidity is within the limits

indicated in § 6.2.1.

 Avoid doing that in rooms where explosive gas, combustible gas, steam or excessive

dust is present.

 Keep you insulated from the object under test.

 Do not touch exposed metal parts such as test lead ends, sockets, fixing objects,

circuits etc.

 Avoid doing that if you notice anomalous conditions such as breakages, deformations,

fractures, leakages of battery liquid, blind display etc.

 Be particularly careful when measuring voltages exceeding 20V to avoid risk of

electrical shocks.


The following symbols are used:

CAUTION - refer to the instruction manual - an improper use may damage the
instrument or its components.

Danger high voltage: risk of electric shocks.

Double insulated meter.

AC voltage or current.

DC voltage or current.


1.1. PRELIMINARY
 This instrument has been designed for use in environments of pollution degree 2.

 It can be used for VOLTAGE and CURRENT measurements on installations of

overvoltage CAT III 1000V and CAT IV 600V.

 When using the instrument always respect the usual safety regulations aimed at

protecting you against the dangerous electric currents and protecting the instrument
against incorrect operations.

 Only the leads supplied with the instrument guarantee compliance with the safety

standards in force. They must be in good conditions and, if necessary, replaced with
identical ones.

 Do not test or connect to any circuit exceeding the specified overload protection.

 Do not effect measurements under environmental conditions exceeding the limits

indicated in § 6.1.1 and § 6.2.1.

 Make sure that batteries are properly installed.

 Before connecting the test probes to the installation make sure that the rotary selector

is positioned on the right function.

 Make sure that LCD and rotary selector indicate the same function