HT-5000

EN - 16

4.3.1.1 Regulating the sensitivity level to match the reception signal strength

Reception signal strength and sensitivity are shown in the following segments of the
display:


Item

Mode of operation

1

Reception signal display bargraph

The strength of the reception signal is represented by this bargraph
scale. These are relative values which are based on the sensitivity
level set in the receiver

The triangles (▲▼) above and below the scale are a visual marker
for maximum values. They move towards the right as long as the
signal strength increases and continue in this direction until it either
levels off or decreases. The maximum marker remains in this position
for 3 seconds and thus provides the user with a helpful visual aid for
determining the maximum signal strength

2

Sensitivity display (gain level in %)
The sensitivity level set in the receiver (which coincides with the total
spectrum which the receiver is capable of processing) is represented
by this scale. The higher the level of sensitivity is set, the more
capable the receiver is of detecting weaker signals.
The numeric display above the bars shows the utilisation factor as a
percent

Should the scale for reception signal strength bargraph provide virtually no indication, or
the display is completely utilised, either a manual or automatic adjustment of the sensitivity
level has to be made.

¾

To start an automatic sensitivity level adjustment, the

button has to be pressed

once. The sensitivity level of the receiver will be automatically adjusted so that the
strength of the signal currently being received will be optimally displayed (fill half of the
scale)

¾

To start a manual sensitivity level adjustment, the

button has to be pressed for

about 2 seconds. The following indicator will appear in the display:

¾

Now the sensitivity of the receiver can be decreased by one percent by pressing the

button, or increased by one percent by pressing the

button. This process can be

accelerated by holding down the respective button

¾

To exit the manual sensitivity level adjustment menu, the

button has to be pressed

for two seconds. By pressing the button only short, the depth measurement process will
begin

2

3

1

2

2

1

2