Technical specifications – HT instruments HT8051 User Manual

Page 16

Advertising
background image

HT8051

EN - 15

6. TECHNICAL SPECIFICATIONS

6.1. TECHNICAL

CHARACTERISTICS

Accuracy is calculated as [%reading + (no. of digits) * resolution] at 18°C

 28°C, <75%RH

Measured DC voltage

Range Resolution Accuracy Input impedance

Protection

against

overcharge

0.01

100.00mV

0.01mV

(0.02rdg +4digits)

1M

30VDC

0.001

10.000V

0.001V

Generated DC voltage

Range Resolution Accuracy

Protection

against

overcharge

0.01

100.00mV

0.01mV

(0.02rdg +4digits)

30VDC

0.001

10.000V

0.001V

Measured DC current

Range Resolution Accuracy

Protection

against

overcharge

0.001

24.000mA

0.001mA

(0.02%rdg + 4digits)

max 50mADC

with 100mA integrated fuse

Sampling resistance: 10

Measured DC current with Loop function

Range Resolution Accuracy

Protection

against

overcharge

0.001

24.000mA

0.001mA

(0.02%rdg + 4digits)

max 30mADC

Sampling resistance: 10

Generated DC current (SOUR and SIMU functions)

Range Resolution Accuracy

Percentage

values

Protection

against

overcharge

0.001

24.000mA

0.001mA

(0.02%rdg + 4digits)

0% = 4mA

100% = 20mA
125% = 24mA

max 24mADC

-25.00

 125.00%

0.01%

SOUR mA mode

 maximum allowed load :1k

 @ 20mA

SIMU mA mode

 loop voltage: 24V rated, 28V maximum, 12V minimum

SIMU Mode

 reference parameters

Loop voltage

Generated current

Load resistance

12V 11mA

0.8k

14V 13mA
16V 15mA
18V 17mA
20V 19mA
22V 21mA
24V 23mA
25V 24mA


Loop mode (loop current)

Range Resolution

Protection

against

overcharge

25VDC ±10%

Not specified

30VDC

Advertising