EN - 3

1.2. DURING USE
Always keep to the instructions contained in this manual.

CAUTION

Non compliance with the CAUTIONs and/or the instructions may damage
the tester and/or its components or injure the operator.

• Before changing the switch position, take off the clamp jaw from the tested conductor

or the electrical circuit in order to avoid any accident

• When the clamp is connected to the circuits to be tested, never touch unused terminals

• When testing resistors, do not add voltage. Although there is a protection circuit,

excessive voltage would cause malfunctioning

• Before measuring current, remove the voltage-resistance test leads

• When measuring current, any strong current close to the clamp jaw will affect the

accuracy

• When measuring current, always put the tested conductor in the middle of the clamp

jaw in order to obtain a more accurate reading

• If the reading value or the sign indication remains unchanged during the measurement,

check if the HOLD function is active


1.3. AFTER USE
• Once the measurements are completed, turn the rotary switch to OFF.

• If you expect not to use the clamp for a long time, remove the battery.

1.4. MEASURING (OVERVOLTAGE) CATEGORIES DEFINITIONS
The norm EN 61010-1: Safety requirements for electrical equipment for measurement,
control and laboratory use, Part 1: General requirements, defines what measuring
category, usually called overvoltage category, is. On § 6.7.4: Measuring circuits, it says:

(OMISSIS)

circuits are divided into the following measurement categories:

• Measurement category IV is for measurements performed at the source of the low-

voltage installation.

Examples are electricity meters and measurements on primary overcurrent protection devices
and ripple control units.

• Measurement category III is for measurements performed in the building installation.

Examples are measurements on distribution boards, circuit breakers, wiring, including cables,
bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and equipment for
industrial use and some other equipment, for example, stationary motors with permanent
connection to fixed installation.

• Measurement category II is for measurements performed on circuits directly

connected to the low voltage installation..

Examples are measurements on household appliances, portable tools and similar equipment..

• Measurement category I is for measurements performed on circuits not directly

connected to MAINS.

Examples are measurements on circuits not derived from MAINS, and specially protected
(internal) MAINS-derived circuits. In the latter case, transient stresses are variable; for that
reason, the norm requires that the transient withstand capability of the equipment is made
known to the user.