2.4 Extended Commands
The Multi-PRO 400 Instrument supports the following extended SDI-12 commands.
Name
Command
Response
Set Factory Defaults 
 
 
 
 
Send Data 
 
aXFD! 
 
 
 
 
aD0! 
 
a0011<CR><LF> 
One result is available in 1 second for reading by 
the Send Data command. Restores all settings and 
calibration values to their factory defaults. 
 
a+s<CR><LF> 
s = command status, 1 = command successful, 0 = 
command failed. 
Set Output Sequence 
 
aXOnnn! 
 
aOK<CR><LF> or 
aERROR<CR><LF> 
 
nnn = 1 to 15 parameter characters in the desired 
output order. The default sequence is oTsCLHr. 
 
o = Dissolved Oxygen Concentration 
T = Temperature 
s = Dissolved Oxygen Saturation 
p = Partial Pressure of Oxygen 
 
A = Actual Conductivity 
C = Specific Conductivity 
S = Salinity 
D = Total Dissolved Solids 
R = Resistivity 
d = Density of Water 
 
P = Pressure 
L = Level 
 
H = pH 
h = pH mV 
r = ORP 
Read Output Sequence
aXO!
annn<CR><LF> 
nnn = 1 to 15 parameter characters as described 
for the XO command. 
Set DO Concentration Units 
 
aXoUnnn! 
 
aOK<CR><LF> or 
aERROR<CR><LF> 
 
nnn = the DO concentration units ID 
Three digits are required. 
Read DO Concentration 
Units 
aXoU!
annn<CR><LF> 
nnn = DO concentration units ID 