3.2.1 Normal Mode
This is the initial measurement mode switched to after completion (or cancellation) of warm-up
.
The readings (measured values) will be updated once each second.
In this mode, measured values cannot be stored.
Various operations can be selected from this screen by using the [LEFT], [RIGHT], [UP],
[DOWN], and [ENTER] keys.
[LEFT] / [RIGHT] Key
On pressing the [LEFT] or [RIGHT] key,
the reading (measured value)will be in “hold”,
and 【HOLD】【PRINT】 will be displayed
on the screen as shown in
Screen-3.
The display will be in “hold” under this condition,
and the reading can be printed from
a printer (optional) by pressing the [ENTER] key.
For releasing hold and returning to the
<NORMAL> Mode (
Screen-2),
press the [LEFT] or [RIGHT] key again.
[UP] / [DOWN] Key
Selecting various measurement modes.
[UP] key (* The [DOWN] key can be used to move in the opposite direction.)
<MENU> → <COUNT> → <P.MODE> → <REPEAT> → <MENU> …
[ENTER] Key
Returning to MENU screen. For details, see 3.3 Menu Setting (page 18).
Screen-2 <
NORMAL> Mode
Remaining warm-up time
*Displayed only when
warm-up is cancelled.
If a printer is not connected, or the baud rates of the printer and the CPC are not
consistent, there will be an error message 【PRINT ERR】 shown at the left bottom
of the screen. Please confirm the connection and communication requirements.
For the CPC baud rate setting method and communication requirements, see 3.3.4.2
Printer Setup (page 25).
Caution
Screen-3
<PRINT> Mode