KANOMAX 6501 series Climomaster User Manual

Page 42

Advertising
background image

41

Transmission of Stored Memory

<To Download the Duct Setting>
1. Type “K ”
2. After “AK”, the data will be displayed.
3. All stored duct settings

(1 through 25) will be downloaded.


Output Content
Entry No. (Data Location); Width; Height; Diameter; Units of Duct
Size

<To Download the Page Number>

1. Type “P”.
2. After “AP”, the data will be displayed.
3. Number of data stored pages will be downloaded.

<To Download the Stored Data>
1. Type “T**** ”.

*

Type in the desired page number(4 digits) after “T”.

2. After “AT”, the data will be displayed.

Output Content

■

Air Velocity Mode: Number of Data; Velocity; Temperature; Humidity

■

Flow Rate Mode: Number of Data; Flow Rate; Temperature; Humidity

■

Pressure Mode: Number of Data; 0000000; 0000000; Pressure

*

The data will be displayed in the currently selected units.

NOT in the units the data was saved with.

*

Calculated data will not be downloaded.









■

The format of the date
output is fixed to
Year/Month/Day. This
is not affected by date
setting of the main
body.

Example

Example

AP
P0011

Example

■

”0000000” will be
displayed for
humidity on
MODEL6541,
6542, 6543 and
6561.

■

”0000000” will be
displayed for
temperature and
humidity on
MODEL6551 and
6552

■

Pressure is only
available on
MODEL6501-B0/-C0.