KYORITSU 2009R User Manual
Page 25

̶ 23 ̶
NOTE
● In the PEAK mode, the auto-ranging feature is disabled and
measuring ranges are fixed as follows.
DC/ACA: 0-400.0A
DC/ACV: 0-400.0A
● When a measured value is 9 counts or less, it is corrected to 0
● The Auto-power-off function is disabled in the PEAK mode as
well.
6-9 Average Measurement
●In the Average mode, "AVG" is indicated on the display.
● The display reads a running average of six readings over an
interval of about 2 seconds.
●This mode is available on ACV, DCV, ACA and DCA ranges.
①Set the function selector switch to the desired position.
② Press the mode selector button once to enter from the normal
mode to the Average mode. "
" should be indicated on
the display.
③Follow instructions for ACV, DCV, ACA or DCA measurement.
④ The display shows a running average of six readings over an
interval of about 2 seconds.