Measuring Ranges and Accuracy
●AC Current
3.Specifications
※Refer to Figure 1 for frequency characteristics.
●Effect of External Stray Magnetic Field
10mA AC max. in proximity to a 10mm-dia conductor carrying 100A
AC.
●Effect of Residual Current
10mA AC max. when clamping on two 10mm-dia conductors, each
carrying supply or return 100A AC current.
Ranges
200mA
2A
20A
200A
1000A
0∼199.9mA
0∼1.999A
0∼19.99A
0∼199.9A
0∼500A
501∼1000A
Accuracy
Frequency Range Response
50/60Hz
WIDE
±1.0%rdg±2dgt
(50/60Hz)
±3.0%rdg±2dgt
(40∼1kHz)
±1.5%rdg±2dgt
(50/60Hz)
±3.5%rdg±2dgt
(40∼1kHz)
±1.5%rdg±2dgt
(50/60Hz)
±3.5%rdg±2dgt
(40∼1kHz)
±5%rdg
(50/60Hz)
±10%rdg
(40∼1kHz)
±1.5%rdg±2dgt
±2.0%rdg±2dgt
±2.0%rdg±2dgt
±5.5%rdg
10min.
Time Limit for
Measurements
Continuous