Specifications – KYORITSU 3131A User Manual

Page 6

Advertising
background image

― 4 ―

3. Specifications

Measuring Range and Accuracy (at 23±5°

C, relative humidity 45-75%)

Insulation Resistance Ranges:(IEC 61557-2)

Nominal Output Voltage

250V

500V

1000V

Measuring Ranges
Open-Circuit

Voltage
Short-Circuit

Current

Nominal Current

Accuracy

0 - 100MΩ

250V DC +20% max.

1.3mA approx

1mA DC min.

at 0.25MΩ

±5% of indicated value

at 0.1MΩ- 10MΩ

±0.7% of scale length at ranges other than above ranges

0 - 200MΩ

500V DC +20% max.

1mA DC min.

at 0.5MΩ

±5% of indicated value

at 0.2MΩ- 20MΩ

0 - 400MΩ

1000V DC + 20% max.

1mA DC min.

at 1MΩ

±5% of indicated value

at 0.4MΩ- 40MΩ

Continuity Test (Resistance Test) Ranges:(IEC 61557-4)

Ranges
Measuring Ranges

x Ω

0 - 2Ω

x 10Ω

0 - 20Ω

Open-Circuit Voltage
Measuring Current
Accuracy
Zero Adjustment Range

4 - 9V DC

200mA DC min.

±3% of scale length

0.2Ω min.

Ranges
250V/100MΩ
500V/200MΩ
1000V/400MΩ

Measuring range to keep operating error

Maximum percentage operating error

0.1MΩ - 10MΩ

±30%

0.2MΩ - 20MΩ
0.4MΩ - 40MΩ

Ranges

x10Ω

Measuring range to keep operating error

Maximum percentage operating error

0.2Ω - 2Ω

±30%

2Ω - 20Ω

Operating Error
Insulation Resistance Ranges(IEC 61557-2)

Continuity Test (Resistance Test) Ranges(IEC 61557-4)

The influencing variations used for calculating the operating error are denoted as follows:

Temperature

: 0°

C and 35°

C

Supply voltage

: 6.5V to 10.4V

Position

: Reference position ±90°

※Prior to measurement, apply Zero Adjustment at each position

Advertising