3.Specifications

● Measuring Range and Accuracy (at 23±5°C, relative humidity 45-75%)

○ Insulation Resistance Ranges:(IEC 61557-2)

Normal output
Voltage

250V

500V

1000V

Measuring Ranges

0 - 100M

Ω

0 - 200M

Ω

0 - 400M

Ω

Open -Circuit

Rated test voltage +20%, -0%

Voltage
Normal current

1mA DC +20%, -0%

Short -Circuit

About 1.3mA DC

Current
Accuracy guaranteed

0.1-10M

Ω

0.2-20M

Ω

0.4-40M

Ω

range

Accuracy

±5% rdg at Accuracy guaranteed ranges
±0.7% of scale length at ranges other than above ranges

○ Continuity Test (Resistance Test) Ranges:(IEC 61557-4)

Ranges

3

Ω

500

Ω

Open -Circuit Voltage

About 4.1V DC

Measuring Current

Greater than 200mA

Accuracy

±1.5% of scale length at other measuring ranges

Operating Error

Insulation Resistance Ranges(IEC 61557-2)

Ranges

Measuring range to keep

Maximum percentage

operating error

operating error

250V/100M

Ω

0.1M

Ω

- 10M

Ω

500V/200M

Ω

0.2M

Ω

- 20M

Ω

±30%

1000V/400M

Ω

0.4M

Ω

- 40M

Ω

― 5 ―

2.Features

MODEL-3132A is an analogue tester with five ranges for insulation resistance
measurement and continuity testing (resistance tests) of low voltage
installations.

● Designed to safety standards:

IEC 61557-1 (General requirements for measuring equipment for low

voltage distribution systems)

IEC 61557-2 (Equipment for insulation resistance measurement for low

voltage distribution systems)

IEC 61557-4 (Equipment for resistance tests for low voltage distribution

systems)

● Dust and drip proof construction to IP54.
● Three insulation test ranges : 250V/100M

Ω

,500V/200M

Ω

,1000V/400M

Ω

.

● Two continuity test ranges : 3

Ω

,500

Ω

● AC voltage warning measurement can be made on all ranges without

depressing the test button.

● Easy for battery check.
● When the test button is released any charge stored in the circuit under test

is automatically discharged.

● Remaining electric charges can be observed on AC voltage warning range.
● LIVE circuit audible and visual indication.
● Fuse protected.
● Color coded scales and range switch position for easy reading.
● Uses only 6 x 1.5V battery type R6P,1.5V AA or equivalent.