KYORITSU 4105A User Manual

Page 5

Advertising
background image

Power Source

9V DC : R6P (SUM-3) x6

Overload Protection

Earth resistance ranges : 280V AC/DC (10 seconds)

Earth voltage range : 300V AC/DC (1 minute)

Insulation Resistance

5M

or more at 500V between the electrical circuit and the housing case

Withstand Voltage

3700V AC for one minute between the electrical circuit and the housing

case

Dimensions

105(L) x 158(W) x 70(D) mm

Weight

Approx. 550g

Accessories

M-7095

Test Leads

x 1 set

M-8032

Auxiliary Earth Spikes

x 2

M-7127

Simplified Measurement Probe

x 1 set

(with safety alligator clip and flat test bar)

M-9084

Carrying Case

x 1

Strap Belt

x 1

Instruction Manual

x 1

Battery R6P (SUM-3)

x 6

---- 7 ----

Maximum Operating Error

Operating error (B) is an error obtained within the rated operating

conditions, and calculated with the intrinsic error (A), which is an error of

the instrument used, and the error (Ei) due to variations.

A

: Intrinsic error

E

1

: Variation due to changing the position

E

2

: Variation due to changing the supply voltage

E

3

: Variation due to changing the temperature

E

4

: Variation due to series interference voltage

E

5

: Variation due to resistance of the probes and auxiliary earth

electrode resistance

E

7

: Variation due to changing the system frequency

E

8

: Variation due to changing the system voltage

Range to keep the maximum operating error

Measurement range within which the maximum operating error (

30%)

applies.

20

range :

5-19.99

200

range :

20-199.9

2000

range :

200-1999

Number of Measurement

3300 times or more

( Measure 10

for 5s on 20

range and take a pause for 25s)

Operating Temperature and Humidity

0 - 40°C, relative humidity 85% or less (no condensation)

Storage Temperature and Humidity

-20 - 60°C, relative humidity 75% or less (no condensation)