Specification – KYORITSU 5406A User Manual
Page 9
― 7 ―
4. SPECIFICATION
● Measurement Specification
Function
Rated
voltage
(AC)
Trip Current
Settings
(I⊿n)
Trip Current
Duration
Accuracy
Trip current
Trip Time
1000ms
1000ms
200ms
1000ms
Goes up by 10%
from 20% to 110%
of I⊿n.
300ms×10
−8%∼−2%
0.6%
±4dgt
+2%∼+8%
−10%∼+10%
−8%∼+8%
10/20/30
/200/300
/500mA
(10/20/30
/200/300
/500mA)
+6mA dc
10/20/30
/200/300
/500mA
230V+10%
-15%
50Hz
Ч1/2
Ч1
Ч5
DC TEST
AUTO RAMP
TEST
Instrument dimensions
186Ч167Ч89mm
Instrument weight
800g
Reference conditions
Specifications are based on the following
conditions except where otherwise stated:
1. Ambient temperature: 23 ± 5 ℃
2. Relative humidity: 45% to 75%
3. Position: horizontal
4. AC power source: 230V, 50Hz
5. Altitude: Up to 2000m
Operating temperature
0 to +40℃, relative humidity 80% or less,
no condensation.
Storage temperature
−20 to +60℃, relative humidity 75% or less,
no condensation.