Operating error, Voltage measurement – KYORITSU 6010B User Manual

Page 9

Advertising
background image

Rated Voltage (AC)

Measuring Range (AC)

Accuracy

100-250V 50Hz

100-300V

3% rdg

― 7 ―

To prevent wrong connection of test leads and to maintain safety, the dedicated
terminals used for continuity and insulation tests are automatically covered
when using the terminals for Loop impedance, RCD and Uc tests.

Typical Number of Tests (central tendency for supply voltage up to 8V at R6P)
Continuity Ranges

: Approx. 700 times min. at load 1Ω

Insulation Resistance Ranges

: Approx. 1200 times min. at load 0.5MΩ(500V)

Approx. 900 times min. at load 1MΩ(1000V)

LOOP/RCD/Uc Ranges

: Operational lifetime:5h (In case of continuous

duty)

Operating error

● Operating Errors of Continuity (IEC 61557-4)/Insulation Resistance (IEC 61557-2)

The influencing variations used for calculating the operating error are denoted as follows;
Temperature

: 0℃ and 35℃

Supply voltage

: 8V to 13.8V

● Operating Error of Loop Impedance (IEC 61557-3)

The influencing variations used for calculating the operating error are denoted as follows:

Temperature

: 0℃ and 35℃

Phase angle

: At a phase angle 0°to 18°

System frequency

: 49.5Hz to 50.5Hz

System voltage

: 230V+10%-15%

Supply voltage

: 8V to 13.8V

Voltage Measurement

@ Loop/RCD/Uc Range

Function

Range

20Ω

200Ω

500V

1000V

0.20-19.99Ω

20.0-199.9Ω

0.50 - 199.9MΩ

1.00 - 199.9MΩ

Measuring range to
keep operating error

Maximum percentage
operating error

±30%

Continuity

lnsulation

Resistance

Range

0.4-19.99Ω

100-1999Ω

Measuring range to keep
operating error

Maximum percentage
operating error

±30%

20Ω

2000Ω