KYORITSU 6011A User Manual

Page 10

Advertising
background image

― 8 ―

The influencing variations used for calculating the operating error are denoted as follows;
Temperature: 0 ℃and 35 ℃
Supply voltage : 8V to 13.8V

● Operating error of Loop Impedance(IEC61557-3)

The influencing variations used for calculating the operating error are denoted as follows:
Temperature:0℃ and 35℃
Phase angle :At a phase angle 0゜to 18゜
System frequency:49.5Hz to 50.5Hz
System voltage:230V+10%-15%

● Operating error of trip current (IEC 61557- 6)

The influencing variations used for calculating the operating error are denoted as follows:
Temperature :0 ℃and 35 ℃
Earth electrode resistance :max 20 Ω
System frequency:49.5Hz to 50.5Hz
System voltage :230V+10%- 15%

Instrument dimensions:-

130 X 183 X 100mm

Instrument weight:-

1080g including batteries.

Reference conditions

Specifications are based on the following conditions
except where otherwise stated:-

1.Ambient temperature: 23± 5:
2.Relative humidity 45% to 75%
3.Position: horizontal
4.AC power source 230V, 50Hz

Range

0.4 ~ 19.99Ω
20.0 ~ 199.9Ω
200 ~ 1999Ω

Measuring range to keep
operating error

Maximum percentage
operating error

±30%

20Ω
200Ω
2000Ω

Function

-10% ∼

0%

0% ∼+10%

-10% ∼+10%

Operating error of trip current

Ч1/2
Ч1
Ч5