8 general specifications – KYORITSU 6017 User Manual

Page 8

Advertising
background image

6

3-8 General Specifications

¡

Operating Temperature

¡

Storage Temperature

¡

Response Time

¡

Effect of Temperature

¡

AC Component in Measuring

Terminal Voltage

(Insulation resistance ranges)

¡

Effect of auxiliary earth resist-

ance (Earth resistance ranges)

¡

Effect of earth voltage

(Earth resistance ranges)

¡

Power Source

¡

Maximum Power Consumption

¡

Insulation Resistance

¡

Withstand Voltage

¡

Overload Protection

: 0 to +40:, 80%RH or less, no condensation.
: -10 to +50:, 75%RH or less, no condensation.
: Insulation resistance ranges: Within 3 sec

Earth resistance ranges: Within 4 sec

Insulation resistance measurement on a
capacitive load can result in a longer response
time.

: Variation in the reading at the points below is

±5% or less when temperature is varied from
20: to 0: and to 40:.
(Primary effective measuring ranges)

: Effects of capacitance elements contained in the

object being measured (5µF) ;Within ±10% of
indicated value. Including fluctuations.

: Within ±5% when fluctuation is 0 to 5kΩ

: Within ±5% at 0 to 5V, ±10% at 0 to 10V
(50/60Hz)

When there is interference voltage of frequency
except 50/60Hz, the indication can be over the
tolerance of the operating error.

: Eight 1.5 batteries, type R6P, AA or equivalent.
: 3.5VA
: 100MΩ or greater between the internal circuit

and the enclosure when measured with 1000V
DC.

: 5550V AC(50/60Hz) for 1 minute between the

internal circuit and the enclosure.

: The instrument operates properly after the

voltages shown in the table below is applied for

Function

Voltage

1000V range:1200V AC

Other ranges: 600V AC

250V AC on all ranges

750V AC

250V AC

Insulation Resistance
Range

Earth Resistance

AC Voltage

Earth Voltage