5.2-4 NULL setting (Protective conductor resistance range)

According to below figure, set the range switch to the range for protective
conductor resistance (either 20

Ω

+ or 20

Ω

- is OK). Press NULL button after

short-circuited PE/2mA and RPE -Test lead terminals with RPE-Test Lead.
Make sure that the "

" mark appears on the LCD and indicated value is

"0.00

Ω

" ; Data is corrected.

CAUTION

●

When pressing NULL button, the indicated value is stored as a reference
value. After that, the difference between the reference value (stored
value) and measuring value is indicated on the LCD.

●

The "

" mark is displayed on the LCD while the NULL function is

operating. Each time pressing NULL button, NULL function shall be
turned on and off.

●

NULL function can not be released by changing the range while it is
operating. To release the function, power off the instrument.

●

NULL function works on both 20

Ω

+ range and 20

Ω

- range, therefore,

there is no need to do NULL setting at both ranges.

●

For the measuring method shown in the following clause 6.3 and 6.4,
NULL setting is required also for an optional cord.

NULL

!

RPE -Test lead terminal

PE/2mA terminal

Fig.9

NULL