Lake Monitors FlowStat ES Sensor (Plastic) User Manual

Page 2

Advertising
background image

(2)

ALL VERSIONS WITH 0-5 VDC CIRCUIT

Power requirements:

12-35VDC, 20 mA

Maximum Current:

30 mA DC

Max. Trans. Distance:

<200 feet recommended

Min. Load Resistance:

1000

Flow rate accuracy:

2% of full scale

Repeatability:

0.5% of scale

Resolution:

infinite

Response time:

2 seconds to 90%

ALL VERSIONS WITH PULSE OUTPUT CIRCUIT

Power requirements:

5-24VDC, 20 mA

Maximum Current:

25 mA DC

Max. Trans. Distance:

<200 feet recommended

Flow rate accuracy:

2% of full scale

Linearity:

0.5% of scale

Resolution:

Infinite

Response time:

< 100 mS

ALL VERSIONS WITH RELAY OUTPUT CIRCUIT

Power Requirements

12-35 V

DC

Max. Trans. Distance

< 200 feet recommended

Switch Contact

Form C, 5A max @120 or 240 VAC

Hysteresis

5% of reading max.

(11)

Introduction

This manual is a service guide produced by the manufacturer. The
manual provides specific procedures and/or illustrations for installa-
tion, inspection, cleaning and filtration of all FlowStat and FlowStat ES
flow sensors. When properly followed, these procedures will keep
your FlowStat flowmeter operating dependably for many years.

It is important for operators and maintenance personnel to be safety
conscience when operating or repairing equipment. Developing a thor-
ough knowledge of the precautionary areas and following safe operat-
ing procedures, can prevent equipment damage and/or personal injury.

Before making any repair, read all of the repair procedures to learn
the correct method and all precautions.

General

Lake Monitors' FlowStat flow rate sensors integrates rugged tangen-
tial turbine technology with a precision digital to analog conversion
circuit hermetically encapsulated within the body of the sensor. The
flow transmitter is ideal for measuring flow rates in cooling circuits,
HVAC systems and batching operations.

Additional features are:

simple in-the-field serviceability of its moving components

the integral 4-20mA circuit design eliminates the need for separate
signal conditioning modules

units come factory calibrated to your system requirements

the low impedance 4-20mA circuit can transmit a "clean" signal
over low-cost wire for several thousand feet without degradation

accurately measures flow in both directions.