Compound Commands

Commands that contain more than one command on one line. For example;

RTD_TYPE PT385_100;RTD_TYPE?

Sets the calibrator to RTD type PT385_100 and queries it to verify. It will return:

PT385_100

13.4.2 Character Processing

The data entered into the calibrator is processed as follows:

• ASCII characters are discarded if their decimal equivalent is less than 32 (space),

except 10 (LF) and 13 (CR):

• Data is taken as 7-bit ASCII

• The most significant data bit is ignored.

• Upper or lower case is acceptable.

13.4.3 Response Data Types

The data returned by the calibrator can be divided into four types:

Integer

For most computers and controllers they are decimal numbers ranging from -32768 to
32768. For example:

FAULT? could return 110

Refer to the Error Codes table (Table 8) for more information on error codes.

Floating

Floating numbers have up to 15 significant figures and exponents. For example:

CPRT_COEFA? returns 3.908300E-03

Character Response Data (CRD)

Data returned as keywords. For example:

RTD_TYPE? returns PT385_100

Indefinite ASCII (IAD)

Any ASCII characters followed by a terminator. For example:

*IDN? returns BETA, 330, 250, 1.00