Thermocouple Read and Source (All errors included) (cont.)
Range
Min
Max
Accuracy
L Thermocouple
-200
0.0
0.45°C
0.0
900.0
0.4°C
U Thermocouple
-200
0.0
0.7°C
0.0
400.0
0.45°C
N Thermocouple
-200
0.0
1.1°C
0.0
1300.0
0.6°C
mV Read/Source
-10.00
75.00
0.015% ±.02
mV
RTD Read and Source
Range
Min
Max
Accuracy
Ni120 (672)
-80.0
260.0
0.2°C
Pt100 (385)
-200.0
800.0
0.33°C
Pt100(3926)
-200.0
630.0
0.3°C
Pt100(3916)
-200.0
630.0
0.3°C
Pt200(385)
-200.0
630.0
0.8°C
Pt500(385)
-200.0
500.0
0.3°C
500.0
630.0
0.4°C
Pt1000(385)
-200.0
100.0
0.2°C
100.0
630.0
0.3°C
Cu10
-100.0
250.0
2.2°C
Cu50
-180.0
200.0
0.5°C
Cu100
-180.0
200.0
0.3°C
YSI400
15.00
50.00
0.05°C
22