Wire functions, Safety, Various operating modes – Martel Electronics QM130M User Manual

Page 2: Power supply, Scaling

Advertising
background image

WIRE FUNCTIONS

RED

V+

Positive power supply connection.

BLACK

0V

Negative power supply connection.

WHITE

Vin

Positive measuring input with reference to 0V.

Range from 0 to 1V d.c.

SAFETY

To comply with the Low Voltage Directive (LVD 93/68/EEC), input voltages to the module’s pins must not exceed 60Vdc. The user must

ensure that the incorporation of the unit into the user’s equipment conforms to the relevant sections of BS EN 61010 (Safety Requirements

for Electrical Equipment for Measuring, Control and Laboratory Use).

VARIOUS OPERATING MODES

Specifications liable to change without prior warning

QM 130M Issue 8

September/2001

M.C.

Applies to QM 130M

Measuring a d.c. current.

Rs =

I

1V

5...12Vd.c.

V+

Vin

0V

0V

Load

Rs

Power

Supply

Measuring a voltage in the range

0 to 1V d.c.

0-1Vd.c.

5...12Vd.c.

+

-

V+

Vin

0V

0V

Simultaneous d.c. Voltage and Current Measurement.

Rs =

I

1V

5...12Vd.c.

V+

V+

Vin

Vin

0V

0V

0V

Load

Rs

Rb

Ra

Power

Supply

I

Current

Voltage

V1

V2

+

-

Ra

Rb

Vin

0V

Voltage

Ra=1M

Rb=Ra.

V1-V2

V2

.

+

-

Rs

I

FSD

Vin

0V

Current

Rs=

I

1V

FSD

SCALING

Two external resistors may be used to alter the full scale

reading of the meter - see table for sample values.

Alternatively, use the following formulae to calculate Ra

and Rb. Select the nearest available standard resistor. To

achieve optimum accuracy, use 1% metal film resistors.

Required F.S.R.

Ra

Rb

10V

910k

100k

100V

1M

10k

10mA

0R

100R

100mA

0R

10R

Martel Electronics, Corp. P.O. Box 770 Londonderry, NH 03053