PIN FUNCTIONS

1. DP1

199.9

2. DP2

19.99

Connect to V+ to display required DP.

3. DP3

1.999

4. TEST

Connect to V+ to display segments as illustrated. It should not be operated for more than a few seconds as the d.c. voltage applied

to the LCD may 'burn' the display. This pin is normally at 5V below V+ and is the ground for the digital section of the meter. It can be

used to power external logic up to a maximum of 1mA.

5. V-

Negative power supply connection.

6. V+

Positive power supply connection.

7. IN HI

Positive measuring differential input.

Analogue inputs must be no closer than 1V to either the positive or negative supply.

8. INLO

Negative measuring differential input.

9. COM

The ground for the analogue section of the A/D converter, held actively at 2.8V (nom.) below V+. COM must not be allowed to sink

excessive current (>100 A) by connecting it directly to a higher voltage.

10. RLO

Negative input for reference voltage (can be connected to COM via Link 3).

11. RHI

Positive input for reference voltage (connected via Link 1 to ROH).

12. ROH

Positive output from internal reference.

µ

To comply with the Low Voltage Directive (LVD 93/68/EEC), input voltages to the module’s pins must not exceed 60Vdc. If voltages to the

measuring inputs do exceed 60Vdc, then fit scaling resistors externally to the module. The user must ensure that the incorporation of the

DPM into the user’s equipment conforms to the relevant sections of BS EN 61010 (Safety Requirements for Electrical Equipment for

Measuring, Control and Laboratory Use).

SAFETY

VARIOUS OPERATING MODES

ON-BOARD LINKS: In order to quickly and easily change operating modes for

different applications, the meter has several on-board links. They are designed

to be easily cut (opened) or shorted (soldered).

Do not connect more than one meter to the same power supply if the meters cannot

use the same signal ground. Taking any input beyond the power supply rails will

damage the meter.

Normally

SHORTED

Cut to

OPEN

Normally

OPEN

Solder to

SHORT

Measuring a floating voltage source

of 200mV full scale.

Split supply operation.

Measuring a supply voltage.

(min. 7.5V, max. 14V).

Measuring the ratio of two voltages.

Reading = 1000 V /V
50mV< V <200mV
V <2V .

1

2

2

1

2

Measuring current. Supply MUST

be isolated.

Measuring 4-20mA to read 0-999.

(supply MUST be isolated).

Check Links 2 & 3 are SHORTED.

Check Link 3 is SHORTED.

Check Links 2 & 3 are SHORTED.

Check Link 3 is SHORTED.

Check Links 1 & 4 are OPEN.

Check Link 3 is SHORTED.

10

7

V-

V-

V+

V+

+

+

+

±200mV

0V

1V

min.

+

-

-

-

-

--

-

V

2

±200mV

V

1

6

5

11

REF HI

IN HI

IN HI

IN HI

REF LO

IN LO

IN LO

IN LO

6

6

6

8

8

8

5

5

5

V-

V-

V-

V-

V-

V-

7

7

7

V+

V+

V+

V+

V+

V+

8

R

+

R=0.2

IF.S.R.

IN

HI

IN

LO

IN

HI

IN

LO

V-

V-

V+

V+

SUPPLY

VOLTAGE

8

7

5

6

510K

10K

510K

V-

V+

8

7

5

6

I

IN

6R2

SET

ZERO

100K

5K

9

IN HI

IN LO

COM

I

OUT

V-

V+

14V max

7V min

Specifications liable to change without prior warning