Ranging dpms for voltage & current, Wiring connections connection descriptions – Martel Electronics DPM65 User Manual

Page 2

Advertising
background image

Ranging DPMs for Voltage & Current

VIN

RA**

RB**

I in

RB**

2V

1 Meg

110 k

2 mA

100

20V

1 Meg

10 k

20 mA

10

200V

1 Meg

1 k

200 mA

1

1000V*

10 Meg***

1 k

* When attentuating voltages above 20V care must be used to use components rated for higher voltages and that

proper creepage and clearance distances are used. Refer to UL3111 or IEC 1010.

** Resistors should be 1%, 1/4 wattt with a 100 or 50 ppm temperature coefficient (Note: Recalibration of span

required if accuracy better than ±1.4% is desired.)

*** Use (2) 4.99 Meg in series if 10 Meg is not available.

Wiring Connections

Connection Descriptions

PIN

Description

V +

+5V DPM Power Supply

GND

DPM Power Supply Ground

DP COM

Decimal Point Return

DP1

1XX.X

DP2

1X.XX

DP3

1.XXX

Input (+)

Positive Input Signal

Input (–)

Negative Input Signal

NC

No Connection Required

HOLD

Hold Last Display

1. Floating Input, DPM65: The input common mode range is ±

1Vdc. If INPUT (–) is not directly conected to GND, a 10k
resistor network can be connected as shown to reduce unstable
readings.

2. Single-Ended Input, DPM65S: The 10K resistor network is not

required. The INPUT (–) is internally connected to GND.

Unused pins should be left open.

CAUTION: Damage to the unit can occur if the power source

polarity is reversed, or greater than 6V is applied
between pins 1 & 2.

DPM-65