Scaling, Applications – Martel Electronics V1 User Manual

Page 4

Advertising
background image

L1

L1

L1

Specifications liable to change without prior warning

V 1

Issue 3

January/2002

M.C.

Applies to V 1/2

Page 4 of 4

Page 4 of 4

SCALING

Two resistors Ra and Rb

may be used to alter the

full scale reading (FSR) of

the meter - see table. The

meter will have to be re-

calibrated by adjusting the

calibration potentiometer

on the rear of the module.

+

-

Ra

Rb

INHI

Vin or in

I

INLO

FSR

Ra

Rb

2V

910k

100k

20V

1M

10k

Vin

200V

1M

1k

2000V*

1M

100R

200 A

0R

1k

2mA

0R

100R

Iin

20mA

0R

10R

200mA

0R

1R

µ

Voltage

Current

*Ensure that Ra is rated for high voltage use.

APPLICATIONS

Do not connect more than one meter to the same power supply

if the meters cannot use the same signal ground. Taking any

input beyond the power supply rails will damage the meter.

Normally

SHORTED

Cut to

OPEN

Normally

OPEN

Solder to

SHORT

0V

0V

+7.0 to +12V

+7.0 to +12V

+3.5 to +6.0V

-3.5 to -6.0V

V+

V+

V+

IN HI

IN HI

IN HI

IN LO

IN LO

IN LO

1

1

3

3

2

2

V 1

V 1

V 1

V-

V-

V-

4

4

DP1

DP1

XDP

XDP

6

6

5

1

3

2

4

DP1

XDP

6

5

5

±200mV

±200mV

+

+

I+

I-

-

-

Measuring an input voltage referenced to a floating supply,

i.e. the input voltage and the meter's power supply are

isolated from each other. Check Link L1 is closed.

Measuring a single ended input referenced to supply, i.e.

the input voltage and the meter's power supply share the

same 0V rail. Check Link L1 is open.

Measuring a current referenced to a floating supply, i.e. the

current being measured and the meter's power supply are

isolated from each other. Check Link L1 is closed.

V 1

3½ Digit LCD Voltmeter Module

www.lascarelectronics.com

www.martelmeters.com

page 4 of 4