EXTERNAL SCALING

Two external resistors Ra and
Rb may be used to alter the
full scale reading (FSR) of the
meter - see table. The meter
will have to be re-calibrated
by adjusting the calibration
potentiometer on the rear of
the module.

FSR

Ra

Rb

2V

910k

100k

20V

1M

10k

Vin

200V

1M

1k

2000V*

1M

100R

200 A

0R

1k

2mA

0R

100R

Iin

20mA

0R

10R

200mA

0R

1R

µ

Voltage

Current

+

-

Ra

Rb

IHI

Vin or in

I

ILO

OEM 1B-LED

*Ensure that Ra is rated for high voltage use.

APPLICATIONS

Do not connect more than one meter to the same power supply if the meters cannot use the same signal
ground. Taking any input beyond the power supply rails will damage the meter.

www.lascarelectronics.com

OEM 1B-LED

3½ Digit LED Voltmeter Module

IHI

ILO

0V

0V

+

-

±200mV

V+

Measuring a single ended input
voltage referenced to supply, i.e. the
input voltage and the meter's power
supply share the same 0V rail.

IHI

ILO

DP1

COM

0V

0V

+

-

±200mV

+4.75 to +5.25V

+4.75 to +5.25V

+4.75 to +5.25V

+4.75 to +5.25V

V+

Measuring an input voltage referenced
to a floating supply, i.e. the input
voltage and the meter's power supply
are isolated from each other.
Circuit shows DP1 enabled.

V+

5k

Set Zero

220k

R

IHI

ILO
COM

0V

0V

I

+

I-

4-20mA

current loop

Measuring a 4-20mA loop current.

The meter's power supply must be
isolated from the 4-20mA current
loop.

R= Reading at 20mA

160

IHI

ILO

COM

0V

0V

R

V+

Measuring current.
The meter's supply is isolated
from the current being measured.

R = 0.2

I

FSR

2

1

3

4

9

5

2

1

3

4

5

2

1

3

4

5

2

1

3

4

COM

IHI

+5V

Load

Shunt

R

L

V+

V

A

S

1

ILO

0V

Rs

7805

220nF

470nF

10uF

V

supply

Measuring supply voltage and current to a load.

Rs = 200mV

Display DP1, DP2 or DP3, by connecting to 0V, as required.

S selects between voltage and current measurement.

Ra and Rb shown scaled for 200V

1

FSR

(e.g. 0.1 / 400mW for 2A full scale)

I

FSR

I

Ra=1M

Rb=1k

2

1

3

4

5