Glossary – Metex 8760CLP Total Free Chlorine & pH Analyzer User Manual

Page 46

Advertising
background image

AQUAMETRIX INC.

1-800-742-1413 www.aquametrix.com

46

GLOSSARY

Electrode

Both a sensing and a reference electrode are needed for the analyzer to measure the process. Commonly these are combined

into one and referred to as a combination electrode. The temperature sensor may be built into the electrode as well.

EPROM

Erasable/Programmable Read Only Memory. The EPROM chip holds the program which determines the functioning of

8760CLP analyzer. Replacing the EPROM chip with a chip containing a new or an updated program changes the way the
analyzer functions. The EPROM chip is programmed by the manufacturer.

Free Available Chlorine

The hypochlorous acid form of chlorine; HOCl.

Hysteresis

The reading at which an alarm is turned on is not the same reading at which the alarm is turned off again. This phenomenon is

referred to as the hysteresis.

LED

Light Emitting Diode. LEDs are used as on/off indicators on the front panel of the 8760CLP.

Menu

The series of prompts which determine the layout of the program used by the analyzer.

Microprocessor

An integrated circuit (chip) which executes the program on the EPROM chip and controls all the input/output functions.

Nernst Equation

Equation which relates the voltage signal produced by the electrodes to the pH of the sample. The equation is temperature

dependent.

NC, Normally Closed

NO, Normally Open.

Normally Closed

Each of the alarm contacts can be wired and configured as normally open or normally closed. A circuit which is wired

normally closed will be closed, ie. the external device wired to it is turned on, when the analyzer is not powered.

Normally Open

A circuit which is wired normally open will be open, ie. the external device wired to it is turned off, when the analyzer is not

powered.

On/off Control

Control response in which the contact is either fully on or fully off.

ppm, Parts per Million.
1 ppm = 1 mg/L. Unit of concentration for chlorine measurement.

RAM

Random Access Memory. Memory in a RAM chip can be both written to and read from. The contents of RAM will disappear

as soon as the RAM chip loses power. The RAM chip has a battery backup device which preserves the contents of the RAM

chip for a considerable time even if the analyzer is turned off. All settings are stored in RAM.

TC, Temperature Compensator.

Temperature Compensation

Correction for the influence of temperature on the sensing electrode. The analyzer reads out concentration as if the process

were at 25 °C, regardless of actual solution temperature.

Total Free Available Chlorine

Sum of the hypochlorous acid (HOCl) and hypochlorite ion (OCl

-

) forms of chlorine.

Advertising