MAINTENANCE AND REPAIR

Page 4

CAUTION: Before breaking the tamper-

evident seal on your meter, be sure that

you are in compliance with any regulatory

requirements (if applicable).

Changing the Pulse Rate. When changing the pulse rate, you

must first remove the existing reed switch (see note above).

1. Choose desired pulse rate and switch position.

2. Remove the black rubber plug from the other hole

and move it to the position the reed switch was

removed from.

3. Insert the reed switch assembly into the side of the

register housing in the direction of the dial with desired

pulse rate.

4. Lock in place with rivet and pin.

Recalibration. For meters used for revenue-billing purposes,

some states require periodic calibration checking. This type

of turbine meter is most commonly checked every four years.

Testing may be done by a local mobile meter service or in a

private or municipal meter shop. Changes in calibration should

be made at an authorized meter shop.

Turbine Insert Replacement. The entire turbine insert comes

out as a unit without removing the meter from the pipe. Carefully

noting position and retaining parts for reassembly.

1. Remove the lid and register as described.

2. After relieving pressure inside the meter, remove

the four flange hex-screws and washers.

3. Lift the flange and insert out in one piece, rocking

gently to break the seal.

4. Remove the four retaining ring screws, and lift

the retaining ring and flange off the turbine insert.

5. Remove and replace the large O-ring around the

top opening of the meter body.

6. Reassemble the retaining ring and flange on top of

the new insert. Be sure the alignment notch in the

retaining ring is directly above the screw in the

insert’s top plate.

7. Replace the entire mechanism in the meter body

with the drilled screw head in the same corner as

the security pin.

8. Replace the lid with a twist of the wrist, insert

the security pin in the tiny hole in the retaining ring,

and thread the security wire through the screw

head and the security pin. Affix lead seal and crimp.

Setting Your Pulse Rate

Size

Size

Pulse

Reed Switch

(inches)

(mm) Rate Position

2”

50

3” 80

4”

100

6”

150

8”

200

10

x1

100

x10

100

x10

1,000

x100

100

x10

1,000

x100

100

x10

1,000

x100

1,000

x100

10,000

x1000

The Register (3” & 4” version shown)

Adding Reed Switch. For the contacting head option, WJ

meters use a meter-mounted reed switch to provide a two-

wire dry contact. To add a reed switch in the field, remove

one of the black rubber plugs from the side of the register

housing and slide the reed switch assembly into the hous-

ing in the direction of the dial with the desired pulse rate.

Lock in place with the rivet and pin (see page 2).

0 0 0 0 0 0

0

5

1

2

3

4

6

7

8

9

U.S. GALLONS X1000

80mm
40°C

X 10

0

5

1

2

3

4

6

7

8

9

X 100

12 345678

Retaining ring

Totalizer

Pulse rate dial #1

Cover Hinge

2” = X100

3” & 4” = X1000

6” & 8” = X10000

2” = X10

3” & 4” = X100

6” & 8” = X1000

Pulse rate dial #2

2” = X1

3” & 4” = X10

6” & 8” = X100

Register Removal. Clip and remove the copper security wire

from the security pin. Remove the security pin that holds

the lid in place, twist the lid and lift it off. Lift the register

free from the meter, noting orientation.

Note: If changing the pulse rate, you need to first remove

the reed switch from the position it is in. In doing so, the

small plastic rivet & pin will be damaged and should not

be reused. Have a replacement plastic pin & rivet before

beginning. If you do not have a spare, contact factory for

a replacement (see page 2 for part reference).