Grounding and troubleshooting, Troubleshooting – Seametrics EM 100-102 v.2 User Manual

Page 4

Advertising
background image

CONNECTIONS

RED + 12 to 24 Vdc Power
BLACK - Power

GREEN + Pulse Output Isolated
WHITE - Pulse Output Isolated

ORANGE + 4-20 mA Isolated
BLUE - 4-20 mA Isolated

}

Optional

STANDARD

GROUNDING

IDEAL GROUNDING

WITH METALLIC PIPING

To Earth Ground

To Earth Ground

GROUNDING and TROUBLESHOOTING

Problem

Probable Causes

Try...

Analog output at 0 mA

Analog output is 4 mA

with flow

No pulse output

Flow rate incorrect

4-20 mA loop needs an external power

supply - if missing, add it

Connect to earth ground

Note flow direction arrow, reverse direction

to match

Change output connection

Connect to earth ground

Change flow direction

Check for proper grounding

Select another flow meter

Install meter in vertical position

No loop power

Unit not grounded

Flow reversed

Reversed output connections

Not grounded

Reversed flow

Missing or incorrect ground wire

Fluid conductivity <20 microSiemens/cm

Empty pipe

LT-65200187-C