Troubleshooting, Calibration (continued) – SHIMPO DFS-series User Manual
Page 19

19
Troubleshooting
The following are general checkpoints; please call your local Shimpo representative or contact Shimpo Instruments directly for
further assistance.
The DFS does not turn on:
• Check all electrical components (power source … battery or power cord)
Error Codes are displayed:
• If any error codes appear on the display, turn power OFF and then ON
• If error code remains, see table below:
Calibration (continued)
DISPLAY
CONDITION
ACTION
EE PROM reading error
Press RESET; unit will default to
factory settings
Memory data error
Press RESET to clear memory
Calibration error
Press RESET and repeat calibration
procedure
Memory unit error; occurs if the
Change measurement units or clear
measurement unit (lb, Kg, N) is
memory
changed while entering data into
memory
Overload condition –possible
Remove excessive load; if the display
load cell damage
does not return to normal operation,
send unit in for repair
MODEL
SELECT NUMBER
DISPLAY
CALIBRATION WEIGHT
DFS-0.5 (R)
00
0.2
200g
DFS-1 (R)
01
0.5
500g
DFS-2 (R)
02
1.0
1000g
DFS-5 (R)
03
2.0
2.0kg
DFS-10 (R)
04
5.0
5.0kg
DFS-20 (R)
05
10.0
10.0kg
DFS-50 (R)
06
20.0
20.0kg
DFS-100 (R)
07
50.0
50.0kg
Please call Shimpo for information on additional sensors.