Troubleshooting

The following are general checkpoints; please call your local Shimpo representative or contact Shimpo Instruments directly for

further assistance.

The DFS does not turn on:

• Check all electrical components (power source … battery or power cord)

Error Codes are displayed:

• If any error codes appear on the display, turn power OFF and then ON
• If error code remains, see table below:

Calibration (continued)

DISPLAY

CONDITION

ACTION

EE PROM reading error

Press RESET; unit will default to

factory settings

Memory data error

Press RESET to clear memory

Calibration error

Press RESET and repeat calibration

procedure

Memory unit error; occurs if the

Change measurement units or clear

measurement unit (lb, Kg, N) is

memory

changed while entering data into

memory
Overload condition –possible

Remove excessive load; if the display

load cell damage

does not return to normal operation,

send unit in for repair

MODEL

SELECT NUMBER

DISPLAY

CALIBRATION WEIGHT

DFS-0.5 (R)

00

0.2

200g

DFS-1 (R)

01

0.5

500g

DFS-2 (R)

02

1.0

1000g

DFS-5 (R)

03

2.0

2.0kg

DFS-10 (R)

04

5.0

5.0kg

DFS-20 (R)

05

10.0

10.0kg

DFS-50 (R)

06

20.0

20.0kg

DFS-100 (R)

07

50.0

50.0kg

Please call Shimpo for information on additional sensors.