7.0 Troubleshooting

Stand does not turn on: FIRST INSPECT all parts for physical interference

Check all electrical components:

• Check for correct AC power source, power cord condition, fuse condition, power switch

(should be “ON”), Emergency cut-off switch setting

• I Inspect test stand fi xtures and force gauge for overload conditions. To disengage

emergency switch, carefully press and turn clockwise.

The force gauge mounting platform does not move:

• Check manual limit knobs and adjust accordingly

• Check to see if the full travel range has already been achieved.

• Check test stand for correct operating mode.

• Check Emergency cut-off switch setting.

Force set points are not stopping the test stand (force gauge platform) movement

Check that communication exists between the force gauge and the test stand

Check that correct cable is fi rmly attached.

Check if the proper baud rate for communication is set from the force gauge. Required

baud rate is 19200 baud rate for FGV-X, FGV-XY. Communication with Shimpo DFS force

gauges bypasses the test stand to the PC when using the FGS-50PCABLE