Techne DB-3A User Manual

Page 21

Advertising
background image

D B - 3 a n d D B - 3 A O P E R A T O R ’ S M A N U A L

P A G E 19

Setting the Overtemperature Cut-out

DB-3

In the event of a fault occurring which could cause the heater to remain
on, power is removed from the heater when the block temperature
reaches approximately 110°C.

DB-3A

An adjustable overtemperature cut-out is fitted to these units. As
supplied it is set to remove power from the heater should the block
temperature exceed approximately 210°C. If you need to prevent the
block temperature exceeding a lower value, the overtemperature cut-
out may be re-adjusted downwards. Note that the cut-out should be set
to trip at least 5°C above the maximum desired operating temperature.

To set the overtemperature cut-out, turn the reset button (located on the
back panel of the unit) fully clockwise then:

either:

• Heat the block to the desired cut-out temperature and turn the reset

button anti-clockwise until the unit just cuts out.

or:

• Heat the block to the required maximum operating temperature and

turn the reset button anti-clockwise until the unit just cuts out. Now
turn the reset button an additional small amount clockwise.

When the cut-out operates, the unit stops working and the
overtemperature cut-out indicator lights.

Remove the block from the unit by screwing the extractor tool into the
thermometer hole and lifting the block vertically. Remember that high
temperatures are dangerous; wear insulating gloves and take extreme
care. Wait for the temperature to fall, perhaps by as much as 40°C, and
press the reset button. The unit will now work as usual.