15 SPECIFICATIONS

SPECIFICATIONS

CON 11

CON 110

Conductivity Range

0 to 19.99 µS/cm,
20.0 to 199.9µS/cm,
200 to 1999 µS/cm;
2.00 to 19.99 mS/cm,
20.0 to 199.9 mS/cm

0 to 19.99 µS/cm,
20.0 to 199.9µS/cm,
200 to 1999 µS/cm;
2.00 to 19.99 mS/cm,
20.0 to 199.9 mS/cm

TDS Range

0.00 to 9.99 ppm
10.0 to 99.9 ppm
100 to 999 ppm;
1.00 to 9.99 ppt
10.0 to 99.9 ppt

0.00 to 9.99 ppm
10.0 to 99.9 ppm
100 to 999 ppm;
1.00 to 9.99 ppt
10.0 to 99.9 ppt

Resolution

0.05 % Full Scale

0.05 % Full Scale

Accuracy

±1% F.S. +1 digit

±1% F.S. +1 digit

Temperature Range

-10.0 to 110.0

°C

-10.0 to 110.0

°C

(14.0 to 230

°F)

Temperature
Resolution

°C

0.1

°C 0.1

°C

Temperature
Resolution

°F

Not available

0.1

°F(from 14.0°F to 199.9°F) &

1.0

°F(from 200°F to 230°F)

Temperature Accuracy ± 0.5

°C ±

0.5

°C (±0.9 °F)

Cell Constant, K

0.1, 1.0, 10.0 (Selectable) 0.1,

1.0,

10.0 (Selectable)

Temperature
Compensation

Automatic / Manual (from 0 to
100°C)

Automatic / Manual (from 0 to
100°C)

Temperature
Coefficient

0.0 to 10.0% / ° C

0.0 to 10.0% / ° C

Normalisation
Temperature

20.0 or 25.0°C

20.0 or 25.0°C

TDS Factor

0.40 to 1.00

0.40 to 1.00

Calibration Points
(Automatic)

84.0, 1413µs/cm; 12.88, 111.8
mS/cm

84.0, 1413µs/cm; 12.88, 111.8
mS/cm

Calibration Points
(Manual)

User defined standard / 1 per
range up to 5 points.

User defined standard / 1 per
range up to 5 points.

Single/Multi point
calibration

Selectable

Selectable

Calibration Range

10% to full scale

10% to full scale

Auto-ranging

Yes (manual select available )

Yes (manual select available )

HOLD Function

Yes

Yes

Memory

50 Data Sets

100 Data Sets

Auto Off

Selectable

Selectable

Auto Off Time

20 minutes after last key
operation

20 minutes after last key
operation