2. SAFETY

PRECAUTIONS


WARNING:
1. This instrument is designed to accept

low level signals supplied by standard
RTDs. Under NO circumstances
should the input voltage exceed the
specified 50V RMS.

2. To prevent ignition of a hazardous

atmosphere, batteries must only be
changed in an area known to be non-
hazardous.

CAUTION:
1. Do not use or store this instrument in

microwave ovens or any abnormally
hot or cold areas.

2. Weak batteries should not be left in

the instrument. Dead batteries can
leak and cause damage to unit.

DANGER:
1. Voltages present at the RTD may

also be present at the battery
terminals. Always disconnect the RTD
when changing batteries.