There are two calibration options:
Offset - adjusts at a single point. Offset
calibration can be performed at any
temperature in the offset range of ±5.00
°C or 9.00 °F
Slope – adjusts at two points. The two
calibration points must be at least 20.00
°C (36.00 °F) apart. The 2nd calibration
point should be at a higher value than the
1

st

calibration point.

Use the ▲ or ▼ keys to adjust the value to

match known temperature standard. Press

F1 to accept.