Thermo Fisher Scientific Thermo Scientific Temp 10 JKT Series User Manual

Page 18

Advertising
background image

- 15 -

17. FIELD CALIBRATION LOCKOUT

AND RE-ENABLE

The calibration lockout feature prevents
any field calibration changes. The lockout
remains in effect until a lockout re-
enable has been performed. Use the
following procedures to lockout or re-
enable the field calibration operation.

LOCKOUT PROCEDURE

1. Turn the thermometer off.
2. Simultaneously press and hold the
cal

and the °C/°F keys down and

momentarily press the on/off key.
Continue to hold the cal and °C/°F keys

until the measurement mode is
displayed.

RE-ENABLE PROCEDURE

1. Turn the thermometer off.
2. Simultaneously press and hold the
hold

and the cal keys down and

momentarily press the on/off key.
Continue to press the hold and cal keys
until the measurement mode is displayed.