Thermo Fisher Scientific CyberScan DO 600 User Manual
Page 74
66
Conductivity Settings Page 2 – Hi-Lo Set Points
Figure 28 : Conductivity Settings Page 2 – Hi-Lo Alarm
This page allows you to set alarm limits for the conductivity measurement mode. 
A visual alarm symbol will appear on the screen, if the conductivity value of the 
measurement is outside of the boundaries set by the minimum and maximum 
limits. 
 
Parameter
Description
Factory Default
Alarm Set
Points
Set alarm limits for the conductivity measurement 
mode: 
ENABLE – The measurement screen shows HI or LO 
alarm indicators when the meter reading is above ‘HI 
alarm value’ or below ‘LO alarm value’ specified in HI 
& LO µS parameters (below) 
DISABLE – HI and LO alarms are not shown in the 
LCD 
DISABLE
Hi ms
Specify Hi alarm value here. The Hi alarm occurs 
when the conductivity reading goes above this value 
Available range: Hi µS (specified below) to 2.000 µS 
(This parameter is not editable when ‘ALARM SET 
POINT’ is disabled) 
2.000 µ
S
Lo
µs
Specify Lo alarm value here. The Lo alarm occurs 
when the conductivity reading goes below this value 
Available range:1.00
µS to Hi mS (specified above)
(This parameter is not editable when ‘ALARM SET 
POINT’ is disabled) 
1.000 µs
Note: 1. If pure water coefficient is set to ‘ENABLE’, the meter will apply pure
water compensation automatically below 2 μS and switch back to linear 
compensation for above 2 µS. 
2. If pure water coefficient is set to ‘DISABLE’, Lnr Temp Coefficient applies 
throughout the measurement.