93
9 DATA
MANAGEMENT
Your CyberScan DO 5000 meter is equipped with an RS-232 port for
sending measurement data to a printer. The printer must be capable
of receiving data within the following guidelines.
Baud rates
110 to 38400 bps
Data Bits
7 or 8
Parity
even, odd or none
Stop Bits
1 or 2
The meter and the external device must be configured to match with
respect to these parameters.
A shielded RS-232 cable must be used. The RS-232 connector on
the back panel of the meter is a female DB9 style. Is the external
device has a 25 pin connector, an adapter is required. The meter’s
pin configuration is:
Pin
Description
1
Buffers analog copy of input
2
Send data from meter
3
Receive data to meter
5
Signal
Common
9
Signal
Common
To initiate printing, touch print if in the manual print mode, or set the
print interval for stable readings or at a timed interval. Refer to the
Setup sections, Set Print Interval and Set Print Criteria.