12.12 Print Functions and Options

This function allows the printing of measurements directly to a
peripheral device via the RS232C port of the meter.

1. Set the mode to the parameter that you wish to print, in the MEAS

mode.

2. Press

PRINT key to send the displayed reading. The printer indicator

will flash to show that data is being sent.

3. ‘ERR“ symbol blinks if the cable is not connected properly or setup

does not match PC requirements.

4. Press

CAL/MEAS to exit error situation.


12.12.1 Print Function Options

To access Print function options, press the SHIFT key and then
the PRINT/OPT key. Option 1 will appear.


1st Option: Print on READY

Default is OFF. Setting the option to ON will cause data to be sent to the
RS232C port whenever the READY indicator appears.

1. Press

SHIFT key followed by the PRINT/OPT key.


2. First option - print on READY - appears. Use

S or T to set to ON.


3. Press

YES to confirm and proceed to the next option.




2nd Option: Print at Time Interval

If no interval is set, the “- - - -”display is shown. Enter the desired
time interval using the numeric keys. The meter sends the data to
the RS232C port at the specified time interval. Press YES to
confirm.

1. Press

SHIFT key followed by the PRINT/OPT key.


2. First option - print on READY - appears. Use

S or T to select the

Print on Time Interval option.


3. Press

YES to confirm and use numeric keypad to set time (between

00.01 to 23.59).


4. Press

YES to confirm and meter will then return to the measurement

mode and automatically starts printing data.

Note: Cable connection and meter protocol setup parameters must be
correct before sending data.