Your meter will normalize its conductivity measurements to a standard temperature that
you can select. You can adjust the normalization temperature from 15 to 30
°C (59 to
86
°F). Meter default is 25.0 °C (77 °F).
1. From measurement mode, press SETUP to enter into setup [SETUP] mode.
2. Use
MI/▲ or MR/▼ key to scroll through the SETUP main-menu P4.0 and press
ENTER thrice repeatedly until you can view the sub-menu P4.2 which the upper
display shows “t.nr” on and “P4.2” on the lower display.
3. Press
ENTER key to make change to the displayed value. The upper display
shows the normalization temperature and the lower display shows “t.nr”.
4. Press
MI/▲ or MR/▼ key to set the normalization temperature (in the specific unit
of temperature which is set at SETUP sub-menu P3.2).
5. Press
ENTER to confirm selection and to return to the SETUP main-menu P4.0.
Press CAL/MEAS key to return to measurement [MEAS] mode.