Thermo Fisher Scientific Ion Selective Electrodes Bromide User Manual

Page 18

Advertising
background image

Instruction Manual

Bromide Electrode

18

Low Slope or

standards contaminated

prepare fresh standards

No Slope or incorrectly made

ISA not used

use recommended ISA


standard used as ISA

use ISA


electrode exposed

check section entitled

to complexing agents

Complexation


air bubble on membrane

remove bubble by

re-dipping probe



"Incorrect Answer" incorrect scaling of

plot millivolts on the linear

(but calibration

semi-log paper

axis. On the log axis, be

curve

is

good)

sure

concentration

numbers

within each decade are

increasing with increasing

concentration.


incorrect sign

be sure to note sign of

millivolt

number

correctly


incorrect standards

prepare fresh standards


wrong units used

apply correct correction factor:

10

-3

M = 79.9 ppm as Br

-1


complexing agents

check section entitled

in sample

Complexation