Carbon Dioxide Electrode

18

incorrect internal filling

refill outer body of electrode

solution

using filling solution shipped

with electrode


total sample level

dilute sample of dissolved

species above 1M


electrode in sample

reduce surface area-to-

too long; CO

2

loss

volume ratio, slow down rate

of stirring, avoid high

temperatures


membrane failure

replace membrane

(wet, perforation, discoloration)


samples and standards

allow samples and standards

not at constant

to come to room temperature

temperature

before use


heat generated by

place insulating material

magnetic stirrer

between stirrer and beaker


defective inner body

refer to

Checking the

Electrode Inner Body


electrode exposed to

hold electrode by outer body

air for extended period

and pull on electrode cable.

Internal filling solution will

flow under membrane and

restore electrode response


Low Slope or No Slope

standards contaminated

prepare fresh standards

or incorrectly made

buffer

not

used use

recommended

buffer


standard used as

use buffer

buffer


electrode exposed

hold electrode by outer body

to air for extended

and pull on electrode cable.

period

Internal filling solution will

flow under membrane and

restore electrode response


membrane failure

replace membrane

(wet, perforation, discoloration)


defective inner body

refer to

Checking the

Electrode Inner Body