Thermo Fisher Scientific Ion Selective Electrodes Copper User Manual
Page 23

23
Try to determine the composition of the samples prior to testing
to eliminate a problem before it starts. See Measuring Hints,
Sample Requirements
, and Interferences.
Technique
Be sure that the electrode's limit of detection has not been
exceeded.
Be sure that the analysis method is clearly understood and that
Good Laboratory Practice has been followed.
Refer to the instruction manual again. Read GENERAL PREPARATION
and ELECTRODE CHARACTERISTICS.
TROUBLESHOOTING HINTS
Symptom
Possible Causes
Next Step
Out of Range
defective meter
check meter with
Reading
shorting
strap
(see meter
instruction manual)
defective
check
electrode
electrode
operation
electrode(s) not
unplug electrode(s)
plugged in properly
and reset
reference electrode
be sure reference
not filled
electrode is filled
air bubble on
remove bubble by
membrane
redipping
electrode
electrode(s) not
put electrode(s)