Thermo Fisher Scientific Ion Selective Electrodes Iodide User Manual

Page 16

Advertising
background image

Instruction Manual

Iodide Electrode

16

standard used as ISA

use ISA


electrode exposed

check section entitled

to complexing agents

Complexation


air bubble on membrane

remove bubble by re-dipping

probe


"Incorrect Answer" incorrect scaling

plot millivolts on the linear axis.

(but calibration

of semi-log paper

On the log axis, be sure that

curve is good)

concentration numbers within

each

decade

are

increasing

with

increasing concentration.


incorrect sign

be sure to note sign of the

millivolt

number

correctly


incorrect standards

prepare fresh standards


wrong units used

apply correct conversion factor:

10

-3

M = 127.0 ppm as I

-1


complexing agents

check section entitled

in sample

Complexation